Advanced Maintenance
Replacing Components of the Analyzer
174
ISQ 7000 Hardware Manual
10. Reattach the screws on each side of the exit endcap.
Figure 146. Reattaching the Quadrupole to the Analyzer Tray
Tip You will get better signal if you gently push the exit endcap forward while tightening
the screws.
11. Once the components are correctly attached to the analyzer tray, follow the instructions
in
"Replacing the Analyzer Tray"
12. Restart your system by following the instructions in
Screws
on
page
162.
"Restarting the System"
Exit Endcap
on
page
165.