Thermo Scientific ISQ 7000 Hardware Manual page 211

Mass spectrometers
Table of Contents

Advertisement

Thermo Scientific
7. Remove the two screws holding the ion guide to the ion guide clamp.
Figure 158. Detaching the Ion Guide from the Ion Guide Clamp
Screws
8. Detach the ion guide from the entrance endcap.
9. Replace the entrance endcap.
10. Attach the ion guide clamp and entrance lens to the new entrance endcap.
11. Slide the ion guide through the ion guide clamp and into the entrance endcap.
Figure 159. Reattaching the Ion Guide
Ion Guide Clamp
12. Use a T10 Torxhead screwdriver to attach the screws holding the ion guide to the ion
guide clamp.
Figure 160. Reattaching the Ion Guide to the Ion Guide Clamp
Screws
3
Advanced Maintenance
Replacing Components of the Analyzer
Ion Guide Clamp
Ion Guide
Entrance Endcap
Ion Guide Clamp
Ion Guide
ISQ 7000 Hardware Manual
183