To replace the components of the source exchange tool
1. Use a T20 Torxhead screwdriver to remove the three screws at the end of the black
handle.
Figure 210. Removing Source Exchange Tool Screws
Screws
2. Loosen the set screw at the other end of the black handle.
3. Slide the black handle toward the other end of the tool.
4. Use a T10 Torxhead screwdriver to remove the set screw holding the endcap onto the
interior shaft.
5. Remove the endcap.
Figure 211. Removing the Endcap
Endcap
6. Slide the handle off the rest of the tool. Set it aside.
7. Remove the interior shaft from the barrel.
Figure 212. Removing the Interior Shaft
Set Screw
Barrel
3
Advanced Maintenance
Replacing Components of the Ion Source
Set Screw
Interior Shaft
ISQ 7000 Hardware Manual
219