7. Look inside the endcap and use a T10 Torxhead screwdriver to remove the two screws
holding the ion guide to the ion guide clamp.
Figure 142. Detaching the Ion Guide from the Ion Guide Clamp
Screws
8. Remove and replace the ion guide clamp.
9. Slide the ion guide and entrance endcap back onto the rods and exit endcap.
10. Reattach the ion guide and quadrupole to the analyzer tray. Make sure the quad and ion
guide are fully forward before tightening screws.
11. Reattach the screws on each side of the quadrupole rod endcaps
Figure 143. Reattaching the Quadrupole and Ion Guide from the Analyzer Tray
Tip You will get better signal if you gently push the quadrupole rod endcaps toward one
another while tightening the screws.
12. Attach the wire connecting the quadrupole entrance lens to the source interface board by
screwing in the terminal with a slotted screwdriver.
13. Once the components are correctly attached to the analyzer tray, follow the instructions
in
"Replacing the Analyzer Tray"
14. Restart your system by following the instructions in
Entrance Endcap
Screws
on
page
162.
"Restarting the System"
3
Advanced Maintenance
Replacing Components of the Analyzer
Ion Guide Clamp
Ion Guide
Exit Endcap
on
page
ISQ 7000 Hardware Manual
165.