Thermo Scientific ISQ 7000 Hardware Manual page 131

Mass spectrometers
Table of Contents

Advertisement

Thermo Scientific
15. Use a T20 Torxhead screwdriver to remove the four screws at each corner of the electron
multiplier plate.
Figure 100. Disconnecting the Electron Multiplier Plate
Electron Multiplier
Plate Screws
CAUTION INSTRUMENT DAMAGE: Use clean, lint-free gloves when touching
components inside the vacuum manifold.
ATTENTION DOMMAGES À L'INSTRUMENT : utilisez des gants propres non pelucheux
lorsque vous touchez des composants à l'intérieur du collecteur à vide.
1
Performing Routine Maintenance
Replacing the Electron Multiplier
ISQ 7000 Hardware Manual
103