Replacing the Boards in the ISQ 7000 Mass Spectrometer
324
ISQ 7000 Hardware Manual
4. The RF board is the second board standing on end toward the center of the ISQ 7000
instrument. Use a T20 Torxhead screwdriver to remove the screw holding the board to
the top of the chassis.
Figure 352. Removing the RF Board and Bracket from the Chassis
5. Use your fingers to pry the bracket of the RF board away from the chassis and toward the
left side of the instrument. Then lift it out of the ISQ 7000 instrument.
6. The lens driver board is the first board standing on end in the center of the instrument.
Figure 353. Locating the Lens Driver Board
RF Board
Chassis
Screw
Lens Driver Board
Bracket
Top Manifold
Cover
Bracket