Advanced Maintenance
Replacing Components of the Analyzer
158
ISQ 7000 Hardware Manual
23. Move the large tab on the analyzer tray to the left, moving the ion source away from the
transfer line, and then carefully pull the tray toward you through the manifold door.
CAUTION - INSTRUMENT DAMAGE Do not scratch any of the sealing surfaces. Be
especially careful when the source interface board reaches the o-ring surface.
ATTENTION DOMMAGES À L'INSTRUMENT Ne rayez pas les surfaces d'étanchéité. Soyez
particulièrement prudent lorsque la carte d'interface de la source entre en contact avec la
surface du joint torique.
Figure 126. Sliding the Analyzer Tray Out of the Instrument
24. Set the tray on a clean work surface.
Analyzer Tray
Tab
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents