Thermo Scientific ISQ 7000 Hardware Manual page 190

Mass spectrometers
Table of Contents

Advertisement

3
Advanced Maintenance
Replacing Components of the Analyzer
Replacing the Analyzer Tray
162
ISQ 7000 Hardware Manual
6. Hold the quadrupole in the tray and make sure the quad assembly and the ion guide are
correctly positioned before tightening the screws on the entrance lens endcap
7. Reattach the screws on each side of the exit endcap. Make sure the quadrupole and exit
endcap are fully forward before tightening screws.
Figure 132. Reattaching the Quadrupole and Ion Guide to the Analyzer Tray
8. Attach the wire connecting the quadrupole entrance lens to the source interface board by
screwing in the terminal with a slotted screwdriver.
To replace the analyzer tray
1. Put tray in manifold with transfer line out or pulled away.
a. If the gas tubes are attached to the transfer line, it will not pull as far away from the
path of the tray as would be best. Be extra careful not to scratch the end of
transferline with the tray, because that could cause a leak with the source plug later.
b. Do not screw in the analyzer tray until the transfer line is fully screwed in.
Note The analyzer tray will mount on one alignment pin in the back and the transfer line
on the right. The mixing chamber will rest on the transfer line when installed.
Entrance Endcap
Screws
Exit Endcap
Thermo Scientific