Thermo Scientific ISQ 7000 Hardware Manual page 208

Mass spectrometers
Table of Contents

Advertisement

3
Advanced Maintenance
Replacing Components of the Analyzer
180
ISQ 7000 Hardware Manual
13. Reattach the exit and entrance endcaps and reattach the screws.
Figure 154. Reattaching the Quadrupole and Ion Guide from the Analyzer Tray
Tip You will get better signal if you gently push the endcaps toward one another while
tightening the screws.
14. Attach the wire connecting the quadrupole entrance lens to the source interface board by
screwing in the terminal with a slotted screwdriver.
15. Once the components are correctly attached to the analyzer tray, follow the instructions
in
"Replacing the Analyzer Tray"
16. Restart your system by following the instructions in
Entrance Endcap
Screws
on
page
162.
"Restarting the System"
Exit Endcap
on
page 165