Ion Source Block - Thermo Scientific ISQ 7000 Hardware Manual

Mass spectrometers
Table of Contents

Advertisement

3
Advanced Maintenance
Replacing Components of the Ion Source

Ion Source Block

186
ISQ 7000 Hardware Manual
To replace the ion source block
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
149.
2. Remove the analyzer tray from the instrument by following the instructions in
"Removing the Analyzer Tray from the ISQ 7000 Instrument"
3. Use a T10 Torxhead screwdriver to disconnect the lens heater block grounding strap from
the lens heater block, but leave it attached to the tray.
Figure 163. Disconnecting the Lens Heater Block Grounding Strap
Lens Heater
Block
Screws
"Shutting Down the ISQ 7000 System"
on
page
151.
Lens Heater Block
Grounding Strap
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents