Advanced Maintenance
Replacing Components of the Vacuum Interlock
236
ISQ 7000 Hardware Manual
34. Reconnect the vacuum interlock cable from the connector.
Figure 240. Reconnecting the Vacuum Interlock Cable
35. Restart your system by following the instructions in
Vacuum
Interlock
Cable
Connector
"Restarting the System"
on
Thermo Scientific
page
165.