Performing Routine Maintenance
Cleaning the Ion Source Cartridge
Figure 82. Adding the Filament to the Advanced EI Ion Source Cartridge
6. Align the filament spacer with the filament assembly. Be sure to align the notch on the
filament spacer with the raised edge of the filament assembly. See
Figure
83.
Figure 83. Aligning the Filament Spacer with the Filament Assembly
Thermo Scientific
ISQ 7000 Hardware Manual
85