Advanced Maintenance
Replacing the Covers of the ISQ 7000 Mass Spectrometer

Top Manifold Cover

280
ISQ 7000 Hardware Manual
To replace the top manifold cover
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
149.
Note You do not have to remove the ion source cartridge or the column. However, you
must vent the instrument.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel screw from the interior
front panel. Then slide the top cover panel toward the back of the instrument and lift it
off.
3. Remove and replace the glass manifold cover on the top of the instrument.
Figure 294. Replacing the Top Manifold Cover
4. Reattach the top cover panel and tighten the screw holding it in place.
5. Restart your system by following the instructions in
Top Manifold Cover
"Shutting Down the ISQ 7000 System"
"Restarting the System"
on
page
165.