Thermo Scientific ISQ 7000 Hardware Manual page 380

Mass spectrometers
Table of Contents

Advertisement

3
Advanced Maintenance
Replacing the Power Supplies
352
ISQ 7000 Hardware Manual
10. Reconnect the large feedthrough cables.
Figure 394. Reconnecting the Large Feedthrough Cables
11. Reattach the support bracket on the left side of the instrument.
Figure 395. Reattaching the Controller Interface Board Support Bracket
Support Bracket
12. Reattach the top cover and left side panels and tighten the screws.
13. Restart your system by following the instructions in
Large
Electron Multiplier
Plate
"Restarting the System"
on
Thermo Scientific
page
165.