3
Advanced Maintenance
Replacing Components of the Vacuum Interlock
224
ISQ 7000 Hardware Manual
3. Disconnect the vacuum interlock cable from the connector.
Figure 220. Disconnecting the Vacuum Interlock Cable
4. Pinch together the two springs on the left side the vacuum interlock shield and remove it.
Figure 221. Removing the Vacuum Interlock Shield
5. Lift the vacuum interlock shield off the instrument. The notch at the bottom of the track
will go around the handle.
Vacuum
Interlock
Cable
Connector
Springs
Thermo Scientific