6. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
Figure 222. Detaching the Manifold Door
7. Set the manifold door on a clean work surface and remove the two screws holding the
pillow block in place.
Figure 223. Removing the Pillow Block
Pillow Block
Screws
Replacing Components of the Vacuum Interlock
Screws
ISQ 7000 Hardware Manual
3
Advanced Maintenance
225