Quadrupole

Thermo Scientific
To replace the quadrupole:
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
149.
2. Remove the analyzer tray from the instrument by following the instructions in
"Removing the Analyzer Tray from the ISQ 7000 Instrument"
3. Remove the screws from the exit endcap.
Figure 145. Removing the Quadrupole from the Analyzer Tray
4. Remove the entire quadrupole, as well as the exit endcap, from the analyzer tray.
5. Slide the quadrupole rods off the exit endcap.
6. Remove the exit endcap from the rods.
7. Attach the quadrupole wires to the new quadrupole rods.
8. Attach the exit endcap to the rods.
9. Slide the new rods and exit endcap onto the ion guide and entrance endcap on the tray.
3
Advanced Maintenance
Replacing Components of the Analyzer
"Shutting Down the ISQ 7000 System"
on
page
151.
Exit Endcap
Screws
ISQ 7000 Hardware Manual
173