Ion Guide Clamp - Thermo Scientific ISQ 7000 Hardware Manual

Mass spectrometers
Table of Contents

Advertisement

3
Advanced Maintenance
Replacing Components of the Analyzer

Ion Guide Clamp

170
ISQ 7000 Hardware Manual
To replace the ion guide clamp:
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
149.
2. Remove the analyzer tray from the instrument by following the instructions in
"Removing the Analyzer Tray from the ISQ 7000 Instrument"
3. Remove the wire connecting the quadrupole entrance lens to the source interface board
by unscrewing the terminal with a slotted screwdriver.
4. Remove two screws on each side of the quadrupole rod endcaps.
Figure 140. Removing the Quadrupole and Ion Guide from the Analyzer Tray
5. Remove the entire quadrupole, as well as the ion guide, from the analyzer tray.
6. Slide the rods and exit quadrupole endcap off the ion guide and entrance endcap.
Figure 141. Separating the Ion Guide from the Quadrupole
"Shutting Down the ISQ 7000 System"
Entrance Endcap
Screws
on
page
151.
Exit Endcap