Performing Routine Maintenance
Cleaning the Ion Source Cartridge
62
ISQ 7000 Hardware Manual
Figure 49. Locating the Ion Source Block
10. Insert the small source removal tool into the ion source block as shown in
Figure 50. Inserting the Small Source Removal Tool
11. Twist the tool to the left. The ion source cartridge should now be connected to the small
source removal tool. See
Figure 51. Twisting the Small Source Removal Tool
12. Slowly remove the ion source cartridge from the instrument.
Figure
51.
Ion Source Block
Figure
50.
Thermo Scientific