Lens Driver Board

Thermo Scientific
To replace the lens driver board
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
149.
Note You do not have to remove the ion source cartridge or the column. However, you
must vent the instrument.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel and left side panel screws
from the interior front panel. Then slide the left and top cover panels toward the back of
the instrument and lift them off.
3. The lens driver board is the first board standing on end in the center of the instrument.
Figure 330. Locating the Lens Driver Board
3
Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
"Shutting Down the ISQ 7000 System"
Lens Driver Board
Bracket
Top Manifold
Cover
ISQ 7000 Hardware Manual
307