Advanced Maintenance
Replacing the Power Supplies
348
ISQ 7000 Hardware Manual
5. Use a T20 Torxhead screwdriver to remove the two screws holding the power supply
system to the top of the back panel.
6. Grasp the power supply system and lift it out of the instrument. It fits snugly, so you may
have to gently shake it as you lift it out.
7. Insert a new power supply and make sure it is flush against the top of the chassis and the
screw holes line up.
8. Reattach the two screws on the back panel.
Tip To make it easier to reattach the top cover panel, pull the sides of the instrument
inward as you tighten the screws.
9. Reconnect the power cables on the back of the instrument. Make sure you attach the
cables properly. The foreline pump connects to the System Mech Pump connector on the
back of the instrument.
10. Reattach the top cover panel and tighten the screw holding it in place.
11. Restart your system by following the instructions in
"Restarting the System"
on
Thermo Scientific
page
165.