Thermo Scientific
5. Twist and remove the tray alignment pins that hold the magnet yoke to the back of the
manifold door. The magnet yoke will pull free of the door, releasing the alignment collar.
Figure 424. Removing the Tray Alignment Pins and Bracket
Magnet Yoke
Tray Alignment
Pins
6. Set the old parts on a clean surface temporarily. There is a place to store them in the
Advanced EI source kit.
7. Remove the new source magnets and magnet yoke from the Advanced EI source kit and
put them together as shown in
Figure 425. Assembling the Advanced EI Source Magnets and Yoke
Magnet Yoke
8. Attach the source magnets and yoke to the manifold door using the new tray alignment
pins in the Advanced EI source kit. See
Figure
425.
Figure
426.
4
Upgrade Equipment
Advanced EI Ion Source
Advanced EI
Source
Magnets
ISQ 7000 Hardware Manual
375