Performing Routine Maintenance
Maintaining the Calibration Gas Module
90
ISQ 7000 Hardware Manual
4. Inject the FC 43. If you see liquid pooled on top of the white frit, remove the excess
liquid according to local environmental regulations.
CAUTION INSTRUMENT DAMAGE: Adding more than 100 μL of calibration compound
at a time can damage the calibration gas module. Be sure liquid does not get into the
module when you reattach the reservoir.
ATTENTION DOMMAGES À L'INSTRUMENT : ajouter plus de 100 μl de composé
d'étalonnage à la fois peut endommager le module de gaz d'étalonnage. Assurez-vous
qu'aucun liquide n'entre dans le module lorsque vous rattachez le réservoir.
5. Reattach the calibrant reservoir and cover to the calibration gas module.
Note For information about replacing the calibrant reservoir, see
Figure 89. Reattaching the Calibrant Reservoir Cover
6. Close the front door of the instrument.
Calibrant Reservoir
Calibration
Gas Module
Calibrant
Reservoir
Cover
Thermo Scientific
Kit.