Thermo Scientific ISQ 7000 Hardware Manual page 366

Mass spectrometers
Table of Contents

Advertisement

3
Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
338
ISQ 7000 Hardware Manual
6. Use your fingers to pry the bracket of the board away from the chassis and toward the left
side of the instrument.
7. Lift the board out of the instrument.
8. Disconnect the Ethernet cable from the PC communication board, which is attached to
the controller interface board.
Figure 374. Disconnecting the Ethernet Cable from the PC Communication Board
Ethernet Cable
Connector
Controller
Interface Board
9. Remove the controller interface board from the instrument.
PC Communication
Board
Thermo Scientific