Thermo Scientific ISQ 7000 Hardware Manual page 414

Mass spectrometers
Table of Contents

Advertisement

4
Upgrade Equipment
CI Reagent Gas Flow Module
386
ISQ 7000 Hardware Manual
15. Reinstall the dynode/multiplier power supply and reattach the screws.
Figure 443. Reinstalling the Dynode/Multiplier Power Supply
Back
Screws
Bottom
Screws
16. Plumb the reagent gas from your gas supply to either port on your CI gas controller.
WARNING FIRE HAZARD: Some CI gases, such as methane and isobutane, are
flammable. Make sure these gases are properly exhausted and all gas fittings on the system
are leak-free. Consult your local Environmental and Safety Regulations for information
about how to properly exhaust fumes from your laboratory.
AVERTISSEMENT RISQUE D'INCENDIE : certains gaz CI, tels que le méthane et
l'isobutane, sont inflammables. Assurez-vous que ces gaz sont correctement évacués et que
tous les raccords de gaz du système sont exempts de fuites. Consultez vos réglementations
locales relatives à l'environnement et à la sécurité pour obtenir des informations sur la
manière d'évacuer correctement les vapeurs de votre laboratoire.
WARNING TOXIC SUBSTANCES HAZARD: Some CI gases, such as ammonia, are toxic.
Make sure these gases are properly exhausted and all gas fittings on the system are
leak-free. Consult your local Environmental and Safety Regulations for information about
how to properly exhaust fumes from your laboratory.