3
Advanced Maintenance
Upgrading the ISQ 7000 System Software

Upgrading Software on the Computer

270
ISQ 7000 Hardware Manual
5. Once the firmware is updated, open the front door of the ISQ 7000 mass spectrometer
and press the Reset button on the interior front panel. This button allows the instrument
to reestablish communication with the computer.
Figure 287. Resetting the System
Reset Button
IMPORTANT These instructions refer to Thermo Foundation 3.1 SP4 and Xcalibur 4.0.
Check the latest Thermo Scientific software CD for the latest versions of Thermo
Foundation and Xcalibur. If the CD contains later versions, follow the steps in the
installation help document on the CD to update your software.
To use Thermo Foundation 3.1 SP4 and Xcalibur 4.0 on Thermo Scientific GC-MS systems,
you must have Xcalibur 4.0 Core Components for GC-MS installed on your computer.
This section describes how to reinstall the Foundation 3.1 SP4 and Xcalibur 4.0 core
components, which are needed to run Foundation 3.1 SP4 and Xcalibur 4.0 on the ISQ 7000
system.
Note There is an electronic version of this installation help document available on the
Xcalibur 4.0Core Components for GC-MS CD. To view this document, select Browse
CD and open the Installation Help.pdf.
Note Before you can load Xcalibur 4.0 Core Components for GC-MS, you must have
administrative privileges on the computer being loaded with the new software.