To replace the calibrant reservoir kit, which consists of the calibrant reservoir and its
cover
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering
these components. If you just need to refill the calibrant reservoir and not replace it, see
"Refilling the Calibrant Reservoir"
1. If you are using hydrogen as a carrier gas, unscrew the hydrogen safety screw on the front
door panel.
2. Open the front door panel of the instrument.
3. Twist the calibrant reservoir cover to disconnect it from the calibration gas module.
Figure 411. Removing the Calibrant Reservoir and Cover
4. Attach a new calibrant reservoir to the calibration gas controller.
Figure 412. Replacing the Calibrant Reservoir and Cover
Calibrant
Reservoir
5. Close the front door panel of the instrument.
6. If you are using hydrogen as a carrier gas, replace the front door panel screw.
Replacing the Calibration Gas Components
on
page
58.
Calibrant
Reservoir
Cover
3
Advanced Maintenance
Calibration
Gas
Calibrant
Reservoir
ISQ 7000 Hardware Manual
363