Thermo Scientific ISQ 7000 Hardware Manual page 348

Mass spectrometers
Table of Contents

Advertisement

3
Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
320
ISQ 7000 Hardware Manual
5. Locate the PC communication board, which is attached to the controller interface board.
Note Although the PC communication board is attached to the controller interface
board, they are sold separately. See the ISQ and TSQ GC-MS Spare Parts Guide for
ordering information.
.
Figure 348. Locating the PC Communication Board
Ethernet Cable
Connector
Controller Interface
Board
6. Disconnect the Ethernet cable from the PC communication board.
7. Remove the controller interface board and PC communication board from the
instrument and replace both or either boards as necessary.
PC Communication
Board