Replacing Components of the Ion Source
200
ISQ 7000 Hardware Manual
8. Reattach the thumbscrew-repeller spacers.
Note When reattaching the spacers, make sure they fit into the grooves on the
thumbscrews. Also, do not over-tighten the thumbscrews or it may crush the spacers.
9. Reattach the source-repeller thumbscrews.
10. Reattach the manifold door and the four screws.
11. Restart your system by following the instructions in
"Restarting the System"
on
Thermo Scientific
page
165.