Maintaining the Foreline Pump
42
ISQ 7000 Hardware Manual
9. Wait five minutes for venting to complete.
CAUTION - INSTRUMENT DAMAGE Do not proceed until the instrument is vented, or
pieces of the column or ferrule might blow into the instrument. To ensure that the
instrument is vented, check how much the glass cover compresses the top cover o-ring in
the manifold. Once the o-ring surface touching the glass is about 1 mm, it is safe to open
the instrument and remove the column.
ATTENTION DOMMAGES À L'INSTRUMENT Ne pas poursuivre tant que l'instrument n'a
pas été ventilé, ou des morceaux de la colonne ou de la virole pourraient éclater dans
l'instrument. Pour vous assurer que l'instrument est ventilé, vérifiez dans quelle mesure le
couvercle en verre comprime le joint torique du couvercle supérieur dans le collecteur.
Lorsque la surface du joint torique en contact avec le verre équivaut à environ 1 mm, vous
pouvez sans danger ouvrir l'instrument et enlever la colonne.
10. Remove one of the oil filler plugs from the foreline pump.
Note The pump oil level must be between the MIN and MAX marks on the oil-level
sight-glass for the pump to operate properly.
11. Add fresh oil to the reservoir until the oil is halfway between the MIN and MAX level
marks.
CAUTION INSTRUMENT DAMAGE: Use only factory-approved foreline pump oil to
avoid damaging the instrument or causing a fire.
ATTENTION DOMMAGES À L'INSTRUMENT : utilisez uniquement de l'huile pour pompe
préliminaire approuvée par le fabricant, afin d'éviter d'endommager l'instrument ou de
provoquer un incendie.
12. If the oil level goes above the MAX level mark, remove the drain plug and drain the excess
oil into a suitable container.
13. Reattach the oil filler plug.
14. Close the front door of the instrument.
15. Twist the vent valve clockwise to close the valve. Be sure not to pinch the vent valve
o-ring.
16. If you are using hydrogen as a carrier gas, replace the front panel screw.
17. Replace all remaining hydrogen safety screws if you are using hydrogen.
18. Power on the ISQ 7000 mass spectrometer.