Replacing The Chassis Feet - Thermo Scientific ISQ 7000 Hardware Manual

Mass spectrometers
Table of Contents

Advertisement

Replacing the Chassis Feet

Thermo Scientific
To replace the chassis feet on the bottom of the ISQ 7000 instrument
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
149.
2. Remove the analyzer tray from the instrument by following the instructions in
"Removing the Analyzer Tray from the ISQ 7000 Instrument"
3. Follow the instructions in the section,
components required to safely replace the chassis feet.
4. Position the instrument so that one foot hangs off the edge of the workbench. You may
have to move the instrument around on the workbench to replace all four feet.
Note If you cannot hang the feet off the edge of the workbench, you can lay the
instrument on its left side and replace the feet. However, you will have to remove the top
cover panel and glass manifold cover first.
5. Use a T20 Torxhead screwdriver to remove the four screws holding any damaged feet to
the bottom of the instrument.
Figure 295. Removing the Feet on the Bottom of the ISQ 7000 Instrument
Feet
Replacing the Covers of the ISQ 7000 Mass Spectrometer
"Shutting Down the ISQ 7000 System"
on
"Distribution Board"
on
ISQ 7000 Hardware Manual
3
Advanced Maintenance
page
151.
page 323
to remove the
281