Performing Routine Maintenance
Cleaning the Ion Source Cartridge
70
ISQ 7000 Hardware Manual
5. With the longer teeth of lens 1 facing down toward lens 2, align the larger metal section
of lens 1 with the lens spring contact slot and let it fall into place.
Figure 62. Inserting Lens 1 into the Ion Cartridge Sleeve
Transfer Line and Lens
Spring Contact Slot
6. Insert the ion volume with the alignment wings fitting into the notches of the sleeve.
Make sure the alignment tooth ion volume is firmly seated into the gap on lens 1.
Figure 63. Inserting the Ion Volume into the Ion Cartridge Sleeve
Larger Metal
Section
Alignment Tooth
Notches
Longer Teeth
of Lens 1
Alignment
Wings
Lens 1 Gap