27. Reattach the screw holding the RF board to the bracket.
Figure 364. Reattaching the RF Board and Bracket
Note This screw is required for good ground contact.
28. Reinstall the lens driver board by aligning the connector on the bottom of the board with
the connectors on the distribution board beneath it. It will drop into the slot when it is in
the right position.
Figure 365. Reinstalling the Lens Driver Board
Connector
3
Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
RF Board
Chassis
Screw
ISQ 7000 Hardware Manual
Bracket
331