Thermo Scientific ISQ 7000 Hardware Manual page 235

Mass spectrometers
Table of Contents

Advertisement

Thermo Scientific
4. Use a T10 Torxhead screwdriver to remove the two screws on the back of the ion source.
Figure 192. Disconnecting the Ion Source from the Analyzer Tray
Screws
5. Slide the source interface board and ion source out from under the tray without damaging
or disconnecting the wires.
Figure 193. Sliding the Ion Source Off the Analyzer Tray
6. Set the source interface board on its side.
Wires
3
Advanced Maintenance
Replacing Components of the Ion Source
ISQ 7000 Hardware Manual
207