4. Look through the left side of the instrument and use a T20 Torxhead screwdriver to
loosen the screw holding the board to the chassis. Loosening the screw will separate the
board from the instrument. The screw is captured, so do not attempt to completely
remove the screw from the board and bracket.
Tip The bracket has a white diagram describing the location of this screw. It is
approximately one inch from the top of the board directly under the white label.
Figure 335. Removing the Lens Driver Board
5. Lift the lens driver board out of the instrument. Be careful as space is limited.
CAUTION - INSTRUMENT DAMAGE Do not touch the components of the boards. Hold
the boards by the sheet metal bracket.
ATTENTION DOMMAGES À L'INSTRUMENT Ne touchez pas les composants des cartes.
Gardez les cartes à proximité du support en tôle.
6. Remove the top manifold cover (glass) out of the top of the instrument and set it aside.
7. Open the front door of the instrument.
3
Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
Lens Driver Board Screw
ISQ 7000 Hardware Manual
311