Exit Endcap - Thermo Scientific ISQ 7000 Hardware Manual

Mass spectrometers
Table of Contents

Advertisement

Exit Endcap

Thermo Scientific
To replace the exit endcap
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
149.
2. Remove the analyzer tray from the instrument by following the instructions in
"Removing the Analyzer Tray from the ISQ 7000 Instrument"
3. Remove the two screws on each side of the exit endcap and slide it off the back of the tray.
Figure 155. Removing the Exit Endcap from the Analyzer Tray
4. Attach a new exit endcap and reattach the two screws.
Tip You will get better signal if you gently push the exit endcap forward while tightening
the screws.
5. Follow the instructions in
6. Restart your system by following the instructions in
"Shutting Down the ISQ 7000 System"
Screws
"Replacing the Analyzer Tray"
3
Advanced Maintenance
Replacing Components of the Analyzer
on
page
151.
Exit Endcap
on
page
162.
"Restarting the System"
on
ISQ 7000 Hardware Manual
page
165.