3
Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
296
ISQ 7000 Hardware Manual
12. Insert a new RF board by aligning the connectors on the bottom of the board with the
connectors on the distribution board beneath it. It will drop into the slot when it is in the
right position.
Figure 314. Reinstalling the RF Board
13. Push the board down into the connectors.
14. Gently push the RF board toward the instrument, carefully aligning the board over the
feedthrough pins, until it stops.
15. Reattach the screw holding the RF board to the bracket.
Figure 315. Reattaching the RF Board and Bracket
Note This screw is required for good ground contact. Be sure it is tightened properly.
RF Board
Chassis
Screw
Bracket
Thermo Scientific