22. Return the glass manifold cover to the top of the instrument.
Figure 96. Replacing the Glass Manifold Cover
Manifold
Door
CAUTION Be sure the glass manifold cover completely covers the o-ring so that the
manifold is properly sealed. The glass cover must not extend forward over the manifold
door.
ATTENTION : assurez-vous que le couvercle du collecteur en verre recouvre complètement
le joint torique afin que le collecteur soit correctement scellé. Le couvercle en verre ne doit
pas dépasser la porte du collecteur.
23. Close the front door of the instrument.
1
Performing Routine Maintenance
Replacing a Dual Filament
O-ring
Glass
Manifold
Cover
ISQ 7000 Hardware Manual
97