11. Flip the manifold door over and remove the o-ring.
Figure 257. Removing the Manifold O-Ring
Manifold
O-ring
12. Flip the door over and twist the vent valve knob and o-ring to remove them.
Figure 258. Removing the Vent Valve Knob and O-Ring
Vent Valve
O-ring
Vent
Valve
Knob
13. Replace the front manifold plate.
Figure 259. Replacing the Front Manifold Plate
14. Insert and tighten the vent valve knob and o-ring into the front manifold plate. Be sure
not to pinch the o-ring.
15. Reattach the front plate o-ring.
• Flip the manifold door over and reattach the magnet yoke, source magnets, and
source retainer magnets.
16. Reattach the three screws and valve body to the front manifold plate.
3
Advanced Maintenance
Replacing Components of the Manifold
ISQ 7000 Hardware Manual
247