Front Door Hinges And Supports - Thermo Scientific ISQ 7000 Hardware Manual

Mass spectrometers
Table of Contents

Advertisement

3
Advanced Maintenance
Replacing Components of the Front Door

Front Door Hinges and Supports

284
ISQ 7000 Hardware Manual
To replace the hinges and supports that connect the front door to the chassis
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
149.
Note You do not have to remove the ion source cartridge or the column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
2. Use a T20 Torxhead screwdriver to loosen the left side panel screw from the interior front
panel.
3. Slide the left side panel toward the back of the instrument and lift it off.
4. Look through the left side of the instrument and remove the nut and grounding washer.
Figure 298. Removing the Nut and Grounding Washer
Grounding Washer
Nut
"Shutting Down the ISQ 7000 System"