Thermo Scientific ISQ 7000 Hardware Manual page 217

Mass spectrometers
Table of Contents

Advertisement

Thermo Scientific
9. Remove the repeller plate and insulating spacers
Figure 168. Removing the Repeller Plate and Insulating Spacers
Insulating
Spacers
10. Use a T10 Torxhead screwdriver to remove the two screws holding the gas mixing
chamber and tubing to the ion source block.
11. Use a T20 Torxhead screwdriver to remove the two screws connecting the lens/source
heater to the ion source block.
Figure 169. Disconnecting the Lens/Source Heater
Replacing Components of the Ion Source
Repeller Plate
ISQ 7000 Hardware Manual
3
Advanced Maintenance
Lens/Source
Heater
Screws
189