Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
332
ISQ 7000 Hardware Manual
29. Push the board down into the connector.
30. Look through the left side of the instrument and use a T20 Torxhead screwdriver to
tighten the screw holding the lens driver board in place on the chassis.
Figure 366. Reattaching the Lens Driver Board
31. Reattach the controller interface board by aligning the connectors on the bottom of the
board with the connectors on the distribution board beneath it. It will drop into the slots
when it is in the right position.
Figure 367. Reinstalling the Controller Interface Board
Connectors
Lens Driver Board Screw