Thermo Scientific ISQ 7000 Hardware Manual page 272

Mass spectrometers
Table of Contents

Advertisement

3
Advanced Maintenance
Replacing Components of the Manifold
244
ISQ 7000 Hardware Manual
3. Disconnect the vacuum interlock cable from the connector.
Figure 251. Disconnecting the Vacuum Interlock Cable
4. Pinch together the two springs on the left side the vacuum interlock shield to remove it.
Figure 252. Removing the Vacuum Interlock Shield
Tubing
Vacuum
Interlock
Cable
Connector
Vacuum
Interlock
Shield
Springs