27. Reconnect the Ethernet cable to the PC communication board.
28. Reinstall the controller interface board by aligning the two connectors on the bottom of
the board to the connectors on the distribution board beneath it. It will drop into the
slots when it is in the right position.
Figure 408. Reinstalling the Control Interface Board
Connectors
29. Reattach the support bracket.
Figure 409. Reattaching the Controller Interface Board Support Bracket
Support Bracket
3
Advanced Maintenance
Replacing the Calibration Gas Components
ISQ 7000 Hardware Manual
361