Thermo Scientific ISQ 7000 Hardware Manual page 241

Mass spectrometers
Table of Contents

Advertisement

Thermo Scientific
4. Use a T20 Torxhead screwdriver to remove the two screws on the back of the ion source.
Figure 201. Disconnecting the Ion Source from the Analyzer Tray
5. Slide the source interface board and ion source out from under the tray without damaging
or disconnecting the wires.
Figure 202. Removing the Ion Source from the Analyzer Tray
Screws
Wires
3
Advanced Maintenance
Replacing Components of the Ion Source
ISQ 7000 Hardware Manual
213