Ion Guide - Thermo Scientific ISQ 7000 Hardware Manual

Mass spectrometers
Table of Contents

Advertisement

Ion Guide

Thermo Scientific
To replace the ion guide:
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
149.
2. Remove the analyzer tray from the instrument by following the instructions in
"Removing the Analyzer Tray from the ISQ 7000 Instrument"
3. Remove the ion source cartridge according to
4. Remove the wire connecting the quadrupole entrance lens to the source interface board
by unscrewing the terminal with a slotted screwdriver.
5. Remove the screws on each side of the endcaps.
Figure 135. Removing the Quadrupole and Ion Guide from the Analyzer Tray
6. Remove the entire quadrupole, as well as the ion guide, from the analyzer tray.
7. Slide the rods and exit endcap off the ion guide and entrance endcap.
Figure 136. Separating the Ion Guide from the Quadrupole
"Shutting Down the ISQ 7000 System"
"Ion Source Cartridge"
Entrance Endcap
Screws
3
Advanced Maintenance
Replacing Components of the Analyzer
on
page
151.
on
page
216.
Exit Endcap
ISQ 7000 Hardware Manual
167