Performing Routine Maintenance
Cleaning the Ion Source Cartridge
56
ISQ 7000 Hardware Manual
13. Pull the vacuum interlock handle down.
Note Be sure to fully extend the tool so that you do not crush the ion volume with the
ball valve.
Figure 40. Pulling the Vacuum Interlock Handle Down
14. Grasp the barrel of the tool and twist it to the left so that the alignment groove moves
toward the left screw. Then pull the tool out of and away from the instrument.
Figure 41. Removing the Tool from the Instrument