Advanced Maintenance
Replacing the Covers of the ISQ 7000 Mass Spectrometer
Right Side Panel
278
ISQ 7000 Hardware Manual
To replace the right side panel of the ISQ 7000 instrument
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
149.
Note You do not have to remove the ion source cartridge. However, you must remove the
column and vent the instrument.
2. If the ISQ 7000 mass spectrometer is too close to the GC, move them apart so you can
access the right side panel.
3. Use a T20 Torxhead screwdriver to remove the screw at each corner of the right side
panel.
Figure 292. Removing the Right Side Panel Screws
"Shutting Down the ISQ 7000 System"
Right Side Panel Screws