8. Reattach the two screws on the back of the ion source.
Figure 205. Reconnecting the Ion Source to the Analyzer Tray
9. Reattach the lens heater block grounding strap to the lens heater block.
Figure 206. Reattaching the Lens Heater Block Grounding Strap
10. Once the components are correctly attached to the analyzer tray, follow the instructions
in
"Replacing the Analyzer Tray"
11. Restart your system by following the instructions in
Screws
Lens Heater Block
Lens Heater Block
Grounding Strap
on
page
162.
3
Advanced Maintenance
Replacing Components of the Ion Source
Screw
"Restarting the System"
on
ISQ 7000 Hardware Manual
page
165.