Thermo Scientific ISQ 7000 Hardware Manual page 99

Mass spectrometers
Table of Contents

Advertisement

Thermo Scientific
7. Insert the large flat end of the repeller so that it rests on top of the ion volume.
Figure 64. Inserting the Repeller into the Ion Cartridge Sleeve
8. Attach the repeller spring to the repeller.
Figure 65. Inserting the Repeller Spring into the Ion Cartridge Sleeve
1
Performing Routine Maintenance
Cleaning the Ion Source Cartridge
Large Flat
End of
Repeller
Repeller
Spring
ISQ 7000 Hardware Manual
71